TT23C three-axis temperature control turntable

Complete technical specifications are listed in the table below. For detailed technical documentation and selection advice, please contact us.

Inquire about this product

Technical Specifications (17)

Load weight≥30kg
Mounting surface diameterΦ520mm
Mounting surface flatness≤0.01mm
Angle measurement accuracy≤±3″
angle measurement repeatability≤±2″
Rate rangeinner axis: 0°/s~1000°/s
middle axis: 0°/s~300°/s
outer axis: 0°/s~300°/s
Rate accuracy and Rate stabilityω<1°/s: ≤1×10⁻³(1° avg)
1<ω<10°/s: ≤1×10⁻⁴(10° avg)
ω>10°/s: ≤1×10⁻⁵(360° avg)
Axis rotary accuracy≤±3″
Axis perpendicularity≤±3″
Axis intersection≤Φ0.5mm sphere inner
Max angular acceleration300°/s²
Temperature range-70℃~120℃
Temperature deviation≤±2℃
Temperature fluctuation≤±0.5℃
Temperature uniformity≤±2℃
Max heating/cooling rate≥3℃/min(at -55℃~+85℃ within the range of linear change, and at 0~3°/min adjustable, requires 30kg aluminum ingot test)
Chamber dimensions≥600mm×600mm×600mm